Timepower TP101 Transformer Oil Dissolved Gas Analysis (DGA) Gas Chromatograph

ASTM D3612 & IEC 60567 Compliant Three-Detector (TCD + Dual FID) Single-Injection Dissolved Gas Analysis System

The Timepower TP101 Gas Chromatograph is engineered for single-injection quantitative determination of dissolved fault gases (H₂, CH₄, C₂H₆, C₂H₄, C₂H₂, CO, CO₂, O₂, N₂) in insulating transformer oils. Equipped with a high-sensitivity TCD, dual FID detectors, nickel methanizer, and built-in IEC/IEEE Duval Triangle fault diagnosis software, it identifies partial discharge, arcing, and thermal faults at an early stage.

Key Technical Features

  • Simultaneous analysis of 7 to 9 dissolved fault gases in insulating oil via a single headspace gas injection
  • Ultra-low detection limit of ≤ 0.1 ppm (µL/L) for Acetylene (C₂H₂) to detect incipient high-energy arcing
  • ±0.05 °C PID column oven temperature stability with automated electronic pressure/flow control
  • Integrated DGA workstation software supporting IEC 60599, IEEE C57.104, Rogers Ratios, and Duval Triangle diagnostics
Models & Technical Specifications
Model / System Measurement Principle & Design Measurement Range & Accuracy Standards & Key Capabilities
TP101 DGA Gas Chromatograph TCD + Dual FID + Nickel Methanizer H₂ ≤ 2 ppm | CO/CO₂ ≤ 2 ppm | Hydrocarbons (C₂H₂) ≤ 0.1 ppm IEC 60567, ASTM D3612, GB/T 17623 (Repeatability RSD ≤ 1.0%)
TP671 Multifunctional DGA Oscillator Automated Headspace Equilibration Shaker Ambient ~ 100 °C (±0.2 °C) | 275 ± 3 rpm, 35 mm stroke Holds 8 × 100 mL syringes, programmable 20 min oscillation + 10 min settling
Applications & Downloadable Catalog
  • Power grid utilities, substations, and transformer preventive maintenance laboratories
  • Transformer manufacturing quality control and high-voltage insulating oil diagnostics
  • Petrochemical gas composition and hydrocarbon impurity testing

📄 Download Timepower Petrochemical & Oil Analyzers Catalog (44 Pages PDF)