HHitech Dura Pro Series Ultrapure Water System (2-Stage RO + Type I)

High-Precision 18.2 MΩ·cm Type I Ultrapure & Pure Water Purification System Compliant with ASTM, CLSI, EP, and USP Standards (Dura Pro 12 / 12F / 12V / 12FV / Dura Pro 24 / 24F / 24V / 24FV)

The HHitech Dura Pro Series Ultrapure Water System (2-Stage RO + Type I) is an advanced laboratory water purification system engineered to deliver consistent Type I ultrapure water (18.2 MΩ·cm @ 25°C) and pure water for analytical, pharmaceutical, biotechnology, and research laboratories. Equipped with DOW RO membranes, nuclear-grade polishing resins, dual-wavelength (185 & 254 nm) UV photo-oxidation, MWCO 5000D ultrafiltration (endotoxin/nuclease removal), and a (0.45+0.1) µm double-layer PES terminal filter, it ensures ultra-low TOC and pyrogen-free water for HPLC, UHPLC, ICP-MS, AAS, PCR, and cell culture applications.

Key Features

  • Output Capacity & Flow Rate: 12 or 24 L/h Production; Up to 2.0 L/min Deionized & Type I Ultrapure Water
  • Control & Monitoring Architecture: 5.0-inch Color Touch Screen (16:9), 3-Way Water Quality Sensors, 2-Way Flow Sensors, 2-Stage RO + 3 Pumps & Built-in Buffer Tank
  • Type I Ultrapure Water Quality: 18.2 MΩ·cm resistivity (25°C), TOC < 3 ppb (UV models) / < 10 ppb (standard), Bacteria < 0.01 CFU/mL, Particles (>0.1 µm) < 1/mL
  • Biological Purity (UF Models): Endotoxin < 0.001 EU/mL, RNases < 1 pg/mL, DNases < 5 pg/mL for sensitive cell culture, IVF, and molecular biology
  • Smart Diagnostics & GLP Compliance: Cartridge life alarm, automatic RO membrane self-flushing, real-time water quality monitoring, and USB/SD data export
Technical Specifications
Parameter / Component Specification
Available Models Dura Pro 12 / 12F / 12V / 12FV / Dura Pro 24 / 24F / 24V / 24FV
Feed Water Requirement Tap Water (5–45°C, 1.0–4.0 kgf/cm²)
Output & Dispensing Rate 12 or 24 L/h Production; Up to 2.0 L/min Deionized & Type I Ultrapure Water
Ultrapure Water Resistivity (25°C) 18.2 MΩ·cm (Meets ASTM, CAP, CLSI, EP, and USP Grade I Standards)
Total Organic Carbon (TOC) < 10 ppb (Standard) / < 3 ppb (Dual-Wavelength 185/254 nm UV Models)
Endotoxin / RNases / DNases < 0.001 EU/mL Endotoxin, < 1 pg/mL RNases, < 5 pg/mL DNases (UF Models)
Bacteria & Particles (>0.1 µm) < 0.01 CFU/mL Bacteria | < 1/mL Particles ((0.45+0.1) µm PES Terminal Filter)
Pure / RO / EDI Water Quality Deionized Water Resistivity > 16 MΩ·cm; 2nd-Stage RO Conductivity 1–5 µS/cm
Control & System Hardware 5.0-inch Color Touch Screen (16:9), 3-Way Water Quality Sensors, 2-Way Flow Sensors, 2-Stage RO + 3 Pumps & Built-in Buffer Tank
Dimensions, Weight & Power 500 × 360 × 540 mm, 26 kg | AC 100–240V, 50/60Hz, 240W
Application Areas
  • Trace analytical instrumentation including HPLC, UHPLC, Ion Chromatography (IC), GC-MS, ICP-MS, ICP-OES, and AAS
  • Molecular biology, PCR, DNA/RNA sequencing, electrophoresis, Mammalian cell culture, and IVF (UF/UVF models)
  • Pharmaceutical QC & R&D laboratories, buffer and standard reagent preparation under GLP/GMP guidelines
  • Feed water for laboratory glassware washers, autoclaves, environmental chambers, and general lab analyzers